Model 650 Conductivity Detector

Access this page in .pdf format & Get Adobe Acrobat[TM] Reader

  • Precise temperature control gives stable, drift-free baselines

  • Detects charged species at ppb levels

  • Compatible with suppressor-based and non­suppressed methods

Unlike other conductivity detectors, the Model 650 can discriminate between random noise and the analyte signal, improving sensitivity. State-of-the-art signal processing and precise temperature control of the detection cell give stable, low-noise baselines. You can easily program and store up to 10 methods through the front panel controls, or control the detector digitally with SSI's EZStart data acquisition and control software.

Detection

Alternating current, synchronous detection, 3 kHz frequency

Range

0.01-5000 µS in 12 steps

Display

Graphical LCD

Autozero

Offsets to 10,000 µS

Time Constant

0.1, 0.5, 1, 5, 10sec, software selectable

Method Storage

0-9, user defined

Output

10mV, 100mV, 1V Full Scale

Computer Interface

RS-232 Serial Communications

Fluid Path

PTFE

Dimensions

10.5" W x 19" L x 5.5" H

 (26.7cm x 48cm x 14cm)

Cell Compartment Size

2.25" W x 6.25" L x 2.25" H

(5.7cm x 15.9cm x 5.7cm)

Insulation

Minimum 1" (2.54cm) all sides

Heater

25W DC heating element vulcanized to

aluminum block

Preheater

PTFE tubing knitted and encased in high

thermal mass housing

Cell Temp.

Factory set at 35°C,

User adjustable ambient to 60°C

Temp. Precision

±0.05°C

Cell Type

2 electrode, gold-plated SS electrodes

Cell Volume

0.5 µL

Pressure Rating

400 psig

RETURN TO TOP OF THE PAGE