home distributors support information
SEARCH | CONTACT US | SECURE SITE

Prep 24/Prep 100 Pump

Access this page in .pdf format & Get Adobe Acrobat[TM] Reader

Preparative Chromatography Pumps

The Prep Pumps have been specifically designed for Preparative Chromatography applications. The units provide high flow rates and high pressure capabilities, with consistent performance to minimize the potential for column damage.

The pumps deliver a maximum flow rate of 24 (6000 psi) or 100 mL/min. (3000 psi).

.Standard controls include RS-232 serial communication port, Run/Stop inputs and remote frequency & voltage controls. Real-time pressure monitoring with user selectable upper & lower pressure limits is standard. All SSI pumps include a 3-year warranty.

Upgrade to High-pressure Gradient:

The combination of two Prep 100s gives you the capability to pump 200 mL/min. at 4000 psi. An optional compact stand-alone keypad is used for convenient high-pressure gradient programming without the requirement for a computer.

Standard Pump Features:

  • Self Flushing Pump Heads and Prime Purge Valve
  • Stainless Steel Fluid Path (PEEK available upon request).
  • Simple Front Panel Key Pad Controls with LED Display (0.1 mL increments)
  • AutoprimeTM one button toggles flow rate to maximize for rapid solvent change
  • Motor Stall Detector
  • Dual ball and seat Check Valves (each head) for consistent flow performance
aaaaaaaaaaa
SPECIFICATIONS  
Flow Rate 0.1-100.0 mL/min; 0.1-24.0 mL/min
Pressure 6000 psi (Prep 24); 3000 psi (Prep 100)
Pulsation ± 1% @ 100 mL/min and 3000 psi
Pressure Accuracy ±2% of full scale pressure
Pressure Zero Offset ±2 psi
Flow Accuracy ±2%
Flow Precision 0.5% RSD
Physical 13.75" W x 17.5" D x 5.5" (26 x 44 x 19 cm) ; 35 lbs (16 kg)

Standard Pump Controls & Options

  • RS-232 Serial Com Port for Complete Control & Status Monitoring
  • Run/stop Inputs (5 volt TTl type)
  • Remote 0-10 volt & remote 0-10 kHz Flow Rate Control input
  • Pressure Control Options: Includes all of the controls above, and real time pressure monitoring with upper/lower pressure limits.
  • High pressure Gradient Control Options: stand-alone keypad unit that enables gradient programming with a second Prep 100.

Gradient System Control Unit Specifications

equilibration time 1-99 min, 1 min increments
equilibration flow rate 0.5-200 mL/min, 0.5 mL/min incr.
equilibration max pressure (Pmax): 0-4000 psi, 10 psi inc.
injection flow rate 0.1-10 mL/min, 0.1 mL/min incr.
injection volume 0.1-99 mL, 0.1 mL incr.
injection delay 0-9 min, 0.1 min incr.
run time: 1-999 min, 1 min incr.
run flow rate: 0.5-200 mL/min, 0.5 mL/min incr.
run P max: 0-4000 psi, 10 psi incr
# of cycles: 1-99
2 COM ports for control of two semi-prep pumps in tandem
1 COM port for Series II pump (for sample introduction)
1 COM port for motorized injection valve Rheodyne PR700-112-02
4 TTL configurable outputs + 4 relays (COM, NC, NO)
dot matrix display (2 lines, 16 characters each), membrane keyboard
configurable from PC-TTL and relay outputs mode (pulse, level logic)

Specifications subject to change without notice.